HUANG KEQIN, XIA YUHUA, and LIU QINGGUO, SEEBECK EFFECT IN OXYGEN PROBE USED FOROXYGEN DETERMINATION IN LIQUID STEEL, J. Univ. Sci. Technol. Beijing, 2(1995), No. 1, pp. 57-62.
Cite this article as:
HUANG KEQIN, XIA YUHUA, and LIU QINGGUO, SEEBECK EFFECT IN OXYGEN PROBE USED FOROXYGEN DETERMINATION IN LIQUID STEEL, J. Univ. Sci. Technol. Beijing, 2(1995), No. 1, pp. 57-62.
HUANG KEQIN, XIA YUHUA, and LIU QINGGUO, SEEBECK EFFECT IN OXYGEN PROBE USED FOROXYGEN DETERMINATION IN LIQUID STEEL, J. Univ. Sci. Technol. Beijing, 2(1995), No. 1, pp. 57-62.
Citation:
HUANG KEQIN, XIA YUHUA, and LIU QINGGUO, SEEBECK EFFECT IN OXYGEN PROBE USED FOROXYGEN DETERMINATION IN LIQUID STEEL, J. Univ. Sci. Technol. Beijing, 2(1995), No. 1, pp. 57-62.
Heat conduction in oxygen probe was analyzed by means of non-steady state method and measured using "double probe" arrangement. The results showed that during the response process the thermal emf was decreased exponentially with the time and in case of low oxygen level the Seebeck coefficient and thermal conductivity of solid electrolyte agreed well with those in literatures.